We compared pre-amplification (PA) RT-PCR blood CD19 mRNA quantification with flow cytometry (FC), to personalize rituximab re-treatment in neuromyelitis optica spectrum disorders (NMOSDs) patients. 47 blood samples from 3 NMOSDs patients were studied. PA-RT-PCR quantified CD19 in all samples, and a positivity threshold was defined, whereas CD19+ B cells were under threshold in 31/47 samples by FC. In all samples where CD19+ B cells were above FC threshold, they resulted above the PA-RT-PCR threshold. CD19 mRNA was above threshold in 8 other samples, resulted negative by FC, and preceded the FC positivity in 7/8 samples by 1-3 months, showing major sensitivity.
Keywords: CD19+ B cells; Flow cytometry; Neuromyelitis optica spectrum disorders; Pre-amplification-real time-PCR; Rituximab re-treatment.